Components of instrumental error of a precision tilt meter, in which pendulum compensation accelerometers are used as sensing elements, are considered. The use of a rotary platform in the design of the tilt meter makes it possible to minimize the effect of temperature on the result of measuring the angle of inclination of the surface, and thus compensate for the change in the zero offset of the accelerometer over time. The errors in measuring the angle of inclination of the foundation on which the test equipment is installed caused by errors in accelerometers and the tilt platform of the tilt meter have been evaluated, and the causes of their occurrence have been investigated. Solutions are proposed to minimize the impact of these errors.