For example,Бобцов

PROBE FOR SCANNING ION-CONDUCTANCE MICROSCOPY

Annotation

  The piezoelectric sensor for scanning probe microscope with a probe in the form of a glass micropipette for scanning ion-conductance microscopy has been made. Resonant frequency and Q-quality of the sensor have been measured in the air and liquid environments. Nanotip has been created for spatial resolution improvement of a microprobe by means of the focused ion beam at a pipette butt end. Using a test grid, the spatial resolution of a microprobe with nanotip is shown. The conclusion about an opportunity of using the sensor with a micropipette in the liquid environment with simultaneous activity in semi contact force mode and in a mode of measurement of ionic currents has been made.

Читать текст статьи

Keywords

Articles in current issue